• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BALT USA BARRICADE COIL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BALT USA BARRICADE COIL SYSTEM Back to Search Results
Model Number 900108
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Balt usa reference (b)(4).Note: this complaint file is being reported late as part of a corrective action following an fda inspection.After a review of historical complaint files, it was determined that for this compaint the device failed to meet its performance specifications according to the initial claim submitted by the user.For these complaints, the potential harm did not lead to a serious injury or death, however the performance specification which was failed to be met for these complaints could have led to a serious injury or death based on the initial information that was reported to balt.The returned device was inspected in our quality laboratory.During our analysis: - we noted that the returned device pouch was opened; - we observed no abnormalities along the pouch seal; - we observed no puncture holes throughout the pouch.Root cause of the packaging issues endured by the coil system can not be definitively determined.Upon return of this device, it was observed that the pouch was already opened.Further investigation showed that there were no visible abnormalities along the pouch seal, indicating proper closure of the aforementioned pouch.Moreover, no puncture holes were seen throughout the pouch.Per lhr, 100% inspection was conducted, with qc final inspection and pouch inspection approval.It is unknown how or when packaging damage occured.Review of the lot history records did not reveal any in-process or lot-specific issue that could account for the observation.No additional complaints against lot number f200300272 has been made for the same issue.Comprehensive analysis of this failure mode has remained subject to monitoring for any unacceptable increase in trend.The submission of this report or related information to the fda, and its release by fda, does not reflect a conclusion by the party submitting this report of the fda that the report or related information is an admission that the manufacturer, their employees, or the device caused or contributed to the reportable event.
 
Event or Problem Description
It was reported that: "at the moment of opening the box containing barricade coil 2x6 the second bag was open so the product was not used because it had lost its sterilization." incident report form received for this complaint indicated no patient injury was sustained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARRICADE COIL SYSTEM
Common Device Name
BARRICADE
Manufacturer (Section D)
BALT USA
29 parker
irvine CA 92618
Manufacturer Contact
david vu
29 parker
irvine, CA 92618
9497881443
MDR Report Key17572782
Report Number3014162263-2023-00032
Device Sequence Number15675168
Product Code HCG
UDI-Device Identifier00818053021077
UDI-Public(01)00818053021077(11)200416(17)250416(10)F200300272
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K151760
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/16/2025
Device Model Number900108
Device Lot NumberF200300272
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2021
Initial Report FDA Received Date08/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-