• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY FLOWGATE2 8F X 85CM; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY FLOWGATE2 8F X 85CM; CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number 90485
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3 other text : the device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during the thrombectomy procedure for a cerebral infarction of left internal carotid artery the subject balloon device was inflated and the thrombectomy was performed.Following this the physician attempted to deflate the subject balloon device but it would not deflate.The subject balloon device was withdrawn to the aortic arch where deflation of the subject balloon device was successfully completed.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Event or Problem Description
It was reported that during the thrombectomy procedure for a cerebral infarction of left internal carotid artery the subject balloon device was inflated and the thrombectomy was performed.Following this the physician attempted to deflate the subject balloon device but it would not deflate.The subject balloon device was withdrawn to the aortic arch where deflation of the subject balloon device was successfully completed.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
There are controls in the manufacturing process to ensure the product met specifications upon release.The device was returned, and reported lot number was not confirmed as the packaging was not returned with the device.During visual inspection, distal part of the balloon catheter was found to be flattened/crushed.The tip of the balloon catheter was found to be damaged.During functional inspection, the balloon could not be inflated due to damage to the device.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint was confirmed based on analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.Additional information received indicated that the device was prepared for use as per the directions for use.There was no damage noted to the packaging prior to opening the packaging and the device was confirmed to be in good condition during preparation/prior to use on the patient.Continuous flush was set up and maintained throughout the clinical procedure and the patients anatomy was described as 'severely tortuous'.During inspection the catheter shaft (distal section) was noted to be damaged (flattened/crushed).The distal tip of the balloon catheter was also damaged.It was not possible to inflate the balloon due to the condition of the returned device.The as reported event of balloon difficult/unable to deflate during use and the as analysed events of balloon failed to inflate, balloon catheter tip damaged and balloon catheter flat/crushed will be assigned procedural factors as this complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural and/or anatomical factors during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOWGATE2 8F X 85CM
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer (Section G)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key17573799
Report Number3012931345-2023-00192
Device Sequence Number15675175
Product Code DQY
UDI-Device Identifier00815742004854
UDI-Public00815742004854
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K131492
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/19/2024
Device Catalogue Number90485
Device Lot Number146064
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/01/2023
Supplement Date Received by Manufacturer10/11/2023
Initial Report FDA Received Date08/18/2023
Supplement Report FDA Received Date10/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-