BRAEMAR MANUFACTURING, LLC EPATCH V2 MB; PATCH IN POUCH ASSY,UNIVERSAL Back to Search Results
Model Number 02-01609
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Skin Tears (2516); Blister (4537); Skin Inflammation/ Irritation (4545)
Type of Reportable Event Serious Injury
Event or Problem Description
The patient reported that after wearing the sensor for 5 days the patient remroved the patch and noted damaged skin with red small dots.A new patch was placed and after 9 days the patient reported she could not moved the patch.After struggling the patch was removed and blood blisters were noted.Sections of the adhesive were noted still on the patient's skin.The patient called her doctor and was prescribed to use an antibotic cream.
 
Additional Manufacturer Narrative
It was reported that the patient experienced skin irritation while wearing the universal patch.The patch did not return for investigation.Engineering evaluation was unable to be performed as the electrode/device was not returned.Allegation is confirmed through images of patient skin irritation and is most probable to be a bio-incompatibility issue with the electrode adhesive.Marsi, skin burn, and associated symptoms may inherently occur under the course of ecg monitoring.No single factor or combination of factors can be attributable to electrode skin irritation and associated symptoms.The product labeling advises patients of alternate options and other steps to take if skin irritation develops, including healthcare professional contact as needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPATCH V2 MB
Common Device Name
PATCH IN POUCH ASSY,UNIVERSAL
Manufacturer (Section D)
BRAEMAR MANUFACTURING, LLC
1285 corporate center drive
suite 150
eagan MN 55121
Manufacturer (Section G)
BRAEMAR MANUFACTURING, LLC
1285 corporate center drive
suite 150
eagan MN 55121
Manufacturer Contact
beverly okoh
1285 corporate center drive
suite 150
eagan, MN 55121
MDR Report Key17573952
Report Number2133409-2023-00047
Device Sequence Number8400306
Product Code DSI
UDI-Device IdentifierB146PE20
UDI-PublicB146PE20
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K081444
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number02-01609
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/08/2023
Supplement Date Received by Manufacturer01/22/2024
Initial Report FDA Received Date08/18/2023
Supplement Report FDA Received Date01/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
Patient SexFemale
-
-