• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ECHELON FLEX 45 COMPACT; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ECHELON FLEX 45 COMPACT; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number SC45A
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 8/18/2023.D4: batch # 730a47.B3: date of event is 2023, event day and month unknown.Captured as (b)(6) 2023.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device visual analysis of the returned sample revealed that one sc45a device was returned with no apparent damage and with no reload present.The device was tested for functionality in the articulated position with a test cartridge reload and achieved its complete firing sequence without any difficulties.The cut line was completed and the staples meet the staple form release criteria.However, the cut was noted to be jagged due to a damaged knife.One possible cause for this type of damage to the knife is when the device is fired over an already existing staple line, hard object or thicker tissue than indicated.Repeatedly firing across existing staple lines can also reduce the ability to cut cleanly.To mitigate the potential for staples getting into the cartridge and interfering with the knife path during device firing, prior to reloading the device, rinse the anvil and cartridge jaw in sterile solution and then wipe the anvil and cartridge jaw to clean any formed but unused staples from the device.Additionally, proper care should be taken when placing the device on the tissue to be stapled, to ensure that no hard obstruction such as a clip is included with the tissue inside the jaws.As part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure that the device meet the required specifications prior to shipment.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 730a47, and no non-conformances were identified.
 
Event or Problem Description
It was reported that after a few stapling, the surgeon felt the knife getting blunt.No patient consequences reported.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON FLEX 45 COMPACT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key17574589
Report Number3005075853-2023-05915
Device Sequence Number12817743
Product Code GDW
UDI-Device Identifier10705036002697
UDI-Public10705036002697
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Catalogue NumberSC45A
Device Lot NumberX96240
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2023
Initial Report FDA Received Date08/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-