MAUDE Adverse Event Report: BALT USA OPTIMA COIL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BALT USA OPTIMA COIL SYSTEM Back to Search Results
Model Number OPTI0103CSS10
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that: "a second optima complex supersoft 1x3 was opened and deployed with ease through the same headway duo 167cm microcatheter.The coil was detached and pusher wire was removed.Dr.(b)(6) noted that as he pulled back mircocatheter that the coil was "stuck" and moving with micro.He introduced a guidewire into microcather and advanced out the tip of catheter manipulating wire back and forth to attempt to dislodge coil.The coil continued to adhere to catheter.Dr.(b)(6) removed microcatheter system and the coil came back with catheter.The physician and tech observed the coil to be "stuck" to outside of micro catheter on back table.A new microcatheter system along with peripheral pushable coils were used to successfully complete case." incident report form submitted for this complaint indicates that no patient injury was sustained.
 
Additional Manufacturer Narrative
Balt usa reference (b)(4).Pending suspected device analysis.The submission of this report or related information to the fda, and its release by fda, does not reflect a conclusion by the party submitting this report of the fda that the report or related information is an admission that the manufacturer, their employees, or the device caused or contributed to the reportable event.
 
Event or Problem Description
It was reported that: "a second optima complex supersoft 1x3 was opened and deployed with ease through the same headway duo 167cm microcatheter.The coil was detached and pusher wire was removed.Dr.(b)(6) noted that as he pulled back mircocatheter that the coil was "stuck" and moving with micro.He introduced a guidewire into microcather and advanced out the tip of catheter manipulating wire back and forth to attempt to dislodge coil.The coil continued to adhere to catheter.Dr.(b)(6) removed microcatheter system and the coil came back with catheter.The physician and tech observed the coil to be "stuck" to outside of micro catheter on back table.A new microcatheter system along with peripheral pushable coils were used to successfully complete case." incident report form submitted for this complaint indicates that no patient injury was sustained.
 
Additional Manufacturer Narrative
Balt usa reference #(b)(4).The returned device was inspected in our quality laboratory.During our analysis: - we noted only the microcatheter and implant coil were returned; - we observed the sr thread on the proximal tip of the implant coil to have a tapered tip at the break point; - we observed the implant coil to be severely damaged near the proximal end; - we observed the microcatheter to have a perforation just proximal of the microcatheter distal tip, with a diameter of approximately 1mm; - we observed the proximal end of the implant coil to be looped within the perforation on the distal tip of the microcatheter; - we observed a significant kink to the lumen of the microcatheter located 62mm proximal of the microcatheter distal tip.Root cause of the reported issue endured by the coil system can be attributed to the implant becoming caught on the damaged microcatheter.Upon device return, we observed the sr thread on the proximal tip of the implant coil to have a tapered tip at the break point as well as the implant coil to be severely damaged near the proximal end.In addition, we observed the microcatheter to have a perforation just proximal of the microcatheter distal tip and the proximal end of the implant coil to be looped within the perforation.We also observed a significant kink to the lumen of the microcatheter located 62mm proximal of the microcatheter distal tip.It was reported a second optima complex supersoft 1x3 was opened and deployed with ease, however, dr.(b)(6) noted that as he pulled back mircocatheter that the coil was "stuck" and moving with microcatheter.From our observations, the implant coil had been threaded through a hole along the wall of the microcatheter at the distal tip.At the perforation, the material from the microcatheter was found to be extruded outward and peeled back along the outer surface.Additional bench testing was performed and indicated the hole found in the returned microcatheter was caused by perforation, and not the detachment of the coil system.Therefore, the reported issue endured by the coil system can be attributed to the implant becoming caught on the damaged microcatheter.The cause of the damage sustained by the microcatheter and the scenario in which the implant ultimately became stuck in the damaged microcatheter cannot be definitively determined or recreated.However, it's likely the perforation was caused from a secondary device.Review of the lot history records did not reveal any in-process or lot-specific issue that could account for the observation.No additional complaints against lot number f220701378 has been made for the same issue.Comprehensive analysis of this failure mode has remained subject to monitoring for any unacceptable increase in trend.The submission of this report or related information to the fda, and its release by fda, does not reflect a conclusion by the party submitting this report of the fda that the report or related information is an admission that the manufacturer, their employees, or the device caused or contributed to the reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIMA COIL SYSTEM
Common Device Name
OPTIMA
Manufacturer (Section D)
BALT USA
29 parker
irvine CA 92618
Manufacturer Contact
david vu
29 parker
irvine, CA 92618
9497881443
MDR Report Key17577610
Report Number3014162263-2023-00051
Device Sequence Number12817760
Product Code HCG
UDI-Device Identifier00818053025723
UDI-Public00818053025723
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K172390
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPTI0103CSS10
Device Lot NumberF220701378
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2023
Supplement Date Received by Manufacturer07/20/2023
Initial Report FDA Received Date08/18/2023
Supplement Report FDA Received Date10/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-