| Catalog Number |
05168589190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The serial number of the c702 analyzer is (b)(6).The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received a questionable result for one patient sample tested with crep2 on a cobas 8000 c702 module.The patient sample initially resulted in a negative value for creatinine on the c702 analyzer.An interference was suspected and a serial dilution of the sample was performed using water and a creatinine result of 30 umol/l was obtained on the c702 analyzer.The patient sample was sent to another laboratory where an unknown creatinine jaffe method was used, resulting in a creatinine value of 56 umol/l.The customer performed a polyethylene glycol (peg) precipitation of the sample and repeated on the c702 analyzer and the creatinine result was 50 umol/l.
|
| |
|
Additional Manufacturer Narrative
|
|
The last calibration performed on (b)(6) 2023 was ok.Quality controls recovered within range.A general reagent issue can be ruled out since calibration and qc data are acceptable.The patient had a high igm result of 17.55 g/l.Per product labeling, "in very rare cases, gammopathy, in particular type igm (waldenström¿s macroglobulinemia), may cause unreliable result." the investigation could not identify a product problem.
|
| |
|