| Medical Device Problem Code |
Fogging (1253)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was foggy image.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Additional Manufacturer Narrative
|
|
His scope was not received for evaluation at stryker endoscopy.This scope was received at henke for evaluation.Based on the henke service record, the reported failure "foggy" was confirmed.According to henke: scope evaluated as a level 3.Distal tip and fiber damaged, broken lenses in system.Probable root cause for this failure is: the complaint for ¿foggy¿ has been confirmed and is due to customer use and handling.Manufacture date is not known.The reported failure mode will be monitored for future reoccurrence.
|
| |
|