• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P 3D KNEE CEMENTED FEMORAL COMPONENT, TINBN COATED, SZ 10, LEFT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P 3D KNEE CEMENTED FEMORAL COMPONENT, TINBN COATED, SZ 10, LEFT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 233-08-110
Medical Device Problem Codes Unstable (1667); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Joint Laxity (4526)
Date of Event 07/24/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery, the agent reported "(instability and infection)".The previous surgery and the surgery detailed in this event occurred 2 months apart.This evaluation is limited in scope as limited information was provided to djo surgical - austin for review.If information regarding cultures identified in the infection, the severity of the infection or any other relevant information is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported devices were the source or had a direct connection with the patient's infection.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was an ncmr#46007 associated with the main part #233-08-110, 3d knee cemented femoral component, tinbn coated, sz 10, left which document that out of 9 parts lot 1 part was rejected due part not coated.Later, the rejected parts were reworked and accepted after proper justification.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to instability and infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.
 
Event or Problem Description
Revision surgery - due to instability and infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3D KNEE CEMENTED FEMORAL COMPONENT, TINBN COATED, SZ 10, LEFT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758
MDR Report Key17578666
Report Number1644408-2023-01099
Device Sequence Number20924841
Product Code JWH
UDI-Device Identifier00888912111362
UDI-Public00888912111362
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K122239
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/12/2024
Device Catalogue Number233-08-110
Device Lot Number452N1917
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2023
Initial Report FDA Received Date08/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/15/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
245-01-110 LOT: 2683A1001; 245-05-011 LOT: 2743A1007; 245-05-011 LOT: 2743A1009; 245-05-111 LOT: 2753A1004; 245-05-111 LOT: 2753A1005; 354-15-025 LOT: 2735A1017; 354-15-025 LOT: 2735A1019
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age59 YR
Patient SexMale
-
-