| Catalog Number |
58-17007E |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported: "the surgeon put a variax 2 1.7 cortex screw through a variax hand plate, and saw on the x-ray that the screw looked too long.After removing the screw, it was noticed that it had uncoiled itself and therefore was too long.".
|
| |
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the received screw was in used condition.The threads of the screw were chipped of in the thread like form.This generally happens when the screw is screwed without proper alignment with the plate and is in direct contact with the plate edge leading to the chip-off of the threads in such manner.The damaged & deformation of the threads on the screw surface also indicates the same.The device history record could not be reviewed because the lot number was not communicated.A review of the labeling did not indicate any abnormalities.Based on the investigation, the root cause was most likely attributed to a user related issue.The failure was caused by mishandling of the device during usage.If any further information is provided, the complaint report will be updated.
|
| |
|
Event or Problem Description
|
|
As reported: "the surgeon put a variax 2 1.7 cortex screw through a variax hand plate, and saw on the x-ray that the screw looked too long.After removing the screw, it was noticed that it had uncoiled itself and therefore was too long.".
|
| |
|