| Model Number |
EB11-J10 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is classified as import for export, therefore 510k is not applicable.Model eb15-j10 is available in the usa with a 510k number k200678.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the operation channel (primary) stuck accessory/object.Based on the result, we concluded that it was caused due to the inadequate/insufficient reprocessing at the facility on the operation channel (primary).Based on the technical report ""hr-rpt-0588(channel)"" and/or the risk analysis results, it was evaluated to submit mdr.
|
| |
|
Event or Problem Description
|
|
The time of event is unknown.There was no report of patient harm.Operation/suction channel clogged.
|
| |
|