| Catalog Number |
623-10-32E |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
07/26/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.H3 other text : not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
Tha for hip osteoarthritis was performed on (b)(6) 2020.The revision surgery was performed on (b)(6) 2023 due to dislocation.The liner and the head were removed in the revision surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding dislocation involving a trident liner was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to dislocation.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
Tha for hip osteoarthritis was performed on (b)(6) 2020.The revision surgery was performed on (b)(6) 2023 due to dislocation.The liner and the head were removed in the revision surgery.
|
| |
|