| Model Number |
N/A |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a hair was found in the packaging of the bearing.There was no delay as the procedure was completed using a second device.There is no additional information available at the time of this report.
|
| |
|
Event or Problem Description
|
|
There is no update to the prior event description provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual evaluation of the provided photo identified a hair-like debris on the device.The sterile packaging has been previously opened; therefore, the source of the debris cannot be confirmed.This complaint cannot be confirmed.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause or condition of the device when it left zimmer biomet cannot be determined, as the product was previously opened.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|