|
Product complaint # (b)(4).The purpose of this mdr submission is to report the findings of the device investigation.Based on the product analysis, the embolic coil of the device received was kinked, the findings meet us regulatory reporting criteria.Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section d2b ¿ procode: krd/hcg a non-sterile galaxy g3 mini 2mmx4mm was received contained in the decontamination pouch.Upon receiving the device, a visual inspection was performed, and it was noted that the coil was inside the introducer.The core wire was found kinked and protruding from the introducer.The coil component was inspected under microscopic magnification, and it was found kinked; this remains attached to the resistance heating (rh) and is still in one piece.The rh was found not softened, indicating that the detachment process was not initiated.The introducer was found opened by a kinked condition at the point where the core wire protrudes.The issue regarding a coil being unable to advance was confirmed since the damages noted in the coil prevented the inability to move the coil.Coil kinking is a known potential issue associated with the use of this device.The instructions for use (ifu) provide proper handling instructions for the device to prevent such issues from occurring.According to the risk documentation, friction and difficulty to advance are potential issues that can occur during microcoil placement due to continuous saline flush not being established, which can result in coil kinking.The damage observed in the coil was not originally documented in the complaint; however, it could be the result of the difficulty experienced during the procedure that could not be replicated in the laboratory.There is no indication that the issue reported is a result of a defect inherently related to the device.The issue regarding a detachable coil delivery system (dcs) being broken was not confirmed since even the dcs was found kinked, this was found in one piece, and the kinked condition is not considered related to the reported issue.This condition may have appeared during the device removal.A manufacturing record evaluation was performed, and no non-conformances related to the reported complaint condition were identified.As part of cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.Devices undergo 100% inspection at final assembly for coil condition to prevent coil damage from leaving the facility.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no capa activity is required.It should be noted that multiple factors could cause product failure.The instructions for use (ifu) does contain the following recommendations: ¿ if unusual friction is noticed during advancement or retraction of the microcoil system through the introducer, open the rhv main valve, and partially withdraw the distal end of the introducer to expose its tip within the rhv.Tighten the rhv main valve and flush the y-connector of the rhv with sterile saline and verify that fluid exits the slit in the clear portion of the introducer.¿ if unusual friction is still noticed during advancement or retraction of the microcoil system, verify flush lines are open and properly pressurized.Then slowly withdraw the entire microcoil system and examine for damage.Replace it with a new microcoil system.If friction still exists, withdraw and examine the delivery catheter system.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
As reported by the field, during a coil embolization, a galaxy g3 mini 2mm x 4cm coil (glm920040, 30742416) could not be advanced from the sheath and subsequently broke from the sheath.After that it could no longer be placed.A new coil with the same dimensions and properties then went smoothly.There was no patient injury reported.No additional information is available.Based on the product analysis of the device received, the embolic coil was kinked.
|