It was reported that on an unknown date, the tip of the cam tightener shaft from the skyline system sheared off in the screw head and could not be removed.The surgeon attempted to remove it with graspers, needles, etc., but was unable to do so.The procedure was completed successfully with no delay.This report involves one skyline anterior cervical plate system cam tightener shaft.This is report 1 of 1 for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: b3: only the event year is known.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|