ALLERGAN (PLEASANTON) COOLSCULPTING SYSTEM; DERMAL COOLING PACK/VACUUM/MASSAGER Back to Search Results
Catalog Number UNK CONTROL UNIT
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hyperplasia (1906)
Type of Reportable Event Serious Injury
Event or Problem Description
Allergan aesthetics received a report of a patient treated with coolsculpting to lower abdomen on (b)(6) 2016 was treated with legacy coolmax applicator who developed paradoxical hyperplasia in abdomen diagnosed on unknown date and month in 2016.
 
Additional Manufacturer Narrative
This is a historical record and reflects reports that were received by the company prior to april 2021.This mdr is associated with a parallel plate applicator, these applicators were discontinued and recalled in 2022.According to the coolsculpting user manual, under rare adverse events, paradoxical adipose hyperplasia (ph/pah) is characterized by a visibly enlarged tissue volume within the treatment area, which may develop two to five months after treatment.Surgical intervention may be required.Pah is not related to any coolsculpting device failure mode but it is included in the risk management files of the device because it is a risk that is inherent to the use cryolipolysis for localized fat reduction.
 
Additional Manufacturer Narrative
Clarification to h.4: disregard previous device manufactured date.
 
Event or Problem Description
Allergan aesthetics received a report of a patient treated with coolsculpting to lower abdomen on (b)(6) 2016 was treated with legacy coolmax applicator who developed paradoxical hyperplasia in abdomen diagnosed on unknown date and month in 2016.
 
Event or Problem Description
Allergan aesthetics received a report of a patient treated with coolsculpting to lower abdomen on (b)(6) 2016 was treated with legacy coolmax applicator who developed paradoxical hyperplasia in abdomen diagnosed on unknown date and month in 2016.
 
Additional Manufacturer Narrative
Corrected data d1 - brand name.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOLSCULPTING SYSTEM
Common Device Name
DERMAL COOLING PACK/VACUUM/MASSAGER
Manufacturer (Section D)
ALLERGAN (PLEASANTON)
4410 rosewood drive
pleasanton CA 94588
Manufacturer (Section G)
ALLERGAN (PLEASANTON)
4410 rosewood drive
pleasanton CA 94588
Manufacturer Contact
terry ingram
4410 rosewood drive
pleasanton, CA 94588
9256214130
MDR Report Key17587516
Report Number3007215625-2023-13356
Device Sequence Number2544594
Product Code OOK
UDI-Device Identifier00816417020230
UDI-Public816417020230
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160259
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberUNK CONTROL UNIT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/06/2017
Supplement Date Received by Manufacturer09/11/2023
10/12/2023
Initial Report FDA Received Date08/21/2023
Supplement Report FDA Received Date09/11/2023
11/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1346-2220
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexFemale
Patient Weight68 KG
Patient RaceBlack Or African American
-
-