| Model Number |
C4120 |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Code |
Bowel Perforation (2668)
|
| Date of Event |
08/02/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Procedure performed: laparascopic right hemicolectomy.Event description: the small bowel was being grasped during hemicolectomy when the event occurred.Message from cns: [name] performing lap right hemi; "yesterday we had a faulty grasper (c4120 ¿ lot:1465118).It had a piece of the mesh material sticking out and it made a small hole in the small bowel.The grasper wasn¿t checked prior to use but was noticed as the surgeon grabbed the bowel with the grasper and a hole formed." the small bowel was punctured.It was subsequently repaired.Device replaced with another c4120.Additional information received from [name] via email on (b)(6) 2023.Small bowel was repaired by suturing.Bleeding and small bowel damage.Intervention: small bowel was repaired by suturing.Device replaced with another c4120.Patient status: small bowel was repaired and patient is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: laparascopic right hemicolectomy.Event description: the small bowel was being grasped during hemicolectomy when the event occurred.Message from cns: [name] performing lap right hemi; "yesterday we had a faulty grasper (c4120 ¿ lot:1465118).It had a piece of the mesh material sticking out and it made a small hole in the small bowel.The grasper wasn¿t checked prior to use but was noticed as the surgeon grabbed the bowel with the grasper and a hole formed.".The small bowel was punctured.It was subsequently repaired.Device replaced with another c4120.Additional information received from [name] via email on 14-aug-2023.Small bowel was repaired by suturing.Bleeding and small bowel damage.Intervention: small bowel was repaired by suturing.Device replaced with another c4120.Patient status: small bowel was repaired and patient is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed a frayed latis pad.Testing was performed on the returned unit; however, the complainant¿s experience of tissue trauma could not be replicated as the latis pads did not snag or puncture tissue during functional testing.Applied medical has reviewed the details surrounding the event and related products and is unable to determine the cause of the reported event based on the evaluation of the returned unit.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
|
| |
|