| Model Number |
101/860/070 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Separation Problem (4043)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the inflation was removed when the tracheostomy tube was replaced.It was not torn.Adverse patient effects are unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: b3: month and year of event have been provided, day is unknown.D4: lot number, expiration date and h4: device manufacture date is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: h3, h6 - updated.One device was returned for evaluation.Under visual inspection it was noted that the inflation line was detached from the pilot balloon.The reported complaint is confirmed.The root cause could not be determined.This issue will continue to be monitored and further actions taken accordingly.A review of the device of the device history records could not be completed as no lot number was provided by the customer.
|
| |
|