• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 367977
Medical Device Problem Codes Inability to Auto-Fill (1044); Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while using bd vacutainer® sst¿ blood collection tubes that there was underfill or low draw and tube push off.This occurred 22 times.The following information was provided by the initial reporter: customer is reporting product 367977 lot no.2306642 unable to draw or stay on the needle.
 
Additional Manufacturer Narrative
Investigation summary: no customer photos or customer samples were received for review and analysis.The device history record was reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.Retain testing cannot be completed due to batch# 2306642 being expired.(product was used by the customer prior to the expiration date).Bd makes no claims on expired product.Bd is unable to confirm the customers reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that while using bd vacutainer® sst¿ blood collection tubes that there was underfill or low draw and tube push off.This occurred 22 times.The following information was provided by the initial reporter: customer is reporting product 367977 lot no.2306642 unable to draw or stay on the needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17591990
Report Number1024879-2023-00589
Device Sequence Number6308393
Product Code JKA
UDI-Device Identifier50382903679770
UDI-Public50382903679770
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
BK050036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2023
Device Catalogue Number367977
Device Lot Number2306642
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2023
Supplement Date Received by Manufacturer09/15/2023
Initial Report FDA Received Date08/21/2023
Supplement Report FDA Received Date09/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-