Model Number 8880828
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a distributor (dis), healthcare provider (hcp) via manufacturer representative regarding spinal product used in posterior spinal fusion.It was reported that when the surgeon has hit the inserter, he felt the rebound of the cage and removed the inserter.He found that the junction of the cage was detached.There were no further complications or symptoms were reported.Additional information was received via manufacturer representative that the reported cage was broken and no fragments left inside the patient.There is no malfunction/allegation associated with the mdt inserter.
 
Event or Problem Description
Additional information was received via manufacturer representative that the reported product was lost.
 
Additional Manufacturer Narrative
B5 has been updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
H3: product analysis of part# 8880828, lot# 0948601w visual and microscopic examination did not reveal any cracks or deformation.The ear/hinge tabs of the implant have been broken off.The damage to the implant appears to be from excessive force placed on the implant during attempted insertion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEVATE¿ SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key17593352
Report Number1030489-2023-00599
Device Sequence Number15704829
Product Code MAX
UDI-Device Identifier00643169430815
UDI-Public00643169430815
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K172199
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8880828
Device Catalogue Number8880828
Device Lot Number0948601W
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2023
Supplement Date Received by Manufacturer10/20/2023
02/21/2024
Initial Report FDA Received Date08/22/2023
Supplement Report FDA Received Date11/14/2023
03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexFemale
-
-