• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA SMS SOLID STEM LAT SIZE 10; HIP STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA SMS SOLID STEM LAT SIZE 10; HIP STEM Back to Search Results
Catalog Number 01.36.070
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 07/25/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 02-aug-2023.Lot 2100584: (b)(4) items manufactured and released on 20-apr-2021.Expiration date: 2026-04-08.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of review.Clinical evaluation performed by medical affairs department: revision 5 days post primary cementless total hip arthroplasty, due to a periprosthetic fracture.It is unknown if a traumatic event occurred.However, weakening of the bone during surgical femoral preparation is a known possible complication of tha.This may favour the occurrence of early fracture during rehabilitation.From the radiographic images, it is visible that a quadra p was implanted and the bone was cabled to fix the fracture.There is no reason to suspect a faulty device.
 
Event or Problem Description
Revision due to periprosthetic bone femoral fracture, at about 5 days post primary.It is unknown if a traumatic event occurred.A quadra stem has been implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMS SOLID STEM LAT SIZE 10
Common Device Name
HIP STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key17594766
Report Number3005180920-2023-00645
Device Sequence Number12617461
Product Code LZO
UDI-Device Identifier07630030888823
UDI-Public07630030888823
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K181693
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/08/2026
Device Catalogue Number01.36.070
Device Lot Number2100584
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2023
Initial Report FDA Received Date08/22/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-