MAUDE Adverse Event Report: ROCHE DIAGNOSTICS IRON2; PHOTOMETRIC METHOD, IRON (NON-HEME)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS IRON2; PHOTOMETRIC METHOD, IRON (NON-HEME) Back to Search Results
Catalog Number 03183696122
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Test Results (2458); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the information provided, the investigation did not identify a product problem.The cause of the event could not be determined.
 
Event or Problem Description
The initial reporter received questionable iron2 iron gen.2 (iron2) results from two patient samples tested on the cobas integra 400 plus.The initial results were not reported outside of the laboratory.Patient sample 1: the patient sample was sent out to another laboratory that uses an au480 analyzer for a rerun.It was also rerun on an alinity analyzer.The initial result from the analyzer was 0.4 umol/l with a data flag.The first repeat result from the analyzer was 0.6 umol/l with a data flag.The second repeat result from the other laboratory's au480 analyzer was 2.6 umol/l.This result was reported outside of the laboratory.The third repeat result from the alinity was 2.0 umol/l.Patient sample 2: the reporter used a new reagent pack (with the same lot number as the previous pack) for the repeat result.The initial result using the previous reagent was 0.8 umol/l.The repeat result using the new reagent pack was 3.1 umol/l.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IRON2
Common Device Name
PHOTOMETRIC METHOD, IRON (NON-HEME)
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17596046
Report Number1823260-2023-02731
Device Sequence Number6430589
Product Code JIY
UDI-Device Identifier04015630918904
UDI-Public04015630918904
Combination Product (Y/N)Y
Initial Reporter CountryBK
PMA/510(K) Number
K951595
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Catalogue Number03183696122
Device Lot Number71147001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2023
Initial Report FDA Received Date08/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-