| Catalog Number |
03183696122 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Based on the information provided, the investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|
Event or Problem Description
|
|
The initial reporter received questionable iron2 iron gen.2 (iron2) results from two patient samples tested on the cobas integra 400 plus.The initial results were not reported outside of the laboratory.Patient sample 1: the patient sample was sent out to another laboratory that uses an au480 analyzer for a rerun.It was also rerun on an alinity analyzer.The initial result from the analyzer was 0.4 umol/l with a data flag.The first repeat result from the analyzer was 0.6 umol/l with a data flag.The second repeat result from the other laboratory's au480 analyzer was 2.6 umol/l.This result was reported outside of the laboratory.The third repeat result from the alinity was 2.0 umol/l.Patient sample 2: the reporter used a new reagent pack (with the same lot number as the previous pack) for the repeat result.The initial result using the previous reagent was 0.8 umol/l.The repeat result using the new reagent pack was 3.1 umol/l.
|
| |
|