ICU MEDICAL COSTA RICA LTD. EXTENSION SET, 15 INCH, 1.2 MICRON FILTER, CLAVE(TM) Y-SITE, SECURE LOCK; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
| Catalog Number |
123540451 |
| Medical Device Problem Code |
Particulates (1451)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.The investigation is still pending.
|
| |
|
Event or Problem Description
|
|
The event involved an extension set, 15 inch, 1.2 micron filter, clave(tm) y-site, secure lock where it was reported that there was condensation inside the tube and that it has white residue.The event occurred during set up (although patient reported this has happened previously as well with dates unknown).Additionally, the product was stored in a controlled temperature/humidity warehouse with continuous monitoring, protected from light.The medication involved was a total parenteral nutrition (tpn).There was no patient harm reported and no delay in therapy.
|
| |
|
Additional Manufacturer Narrative
|
|
Received two used 123540451 extension set, 1.2 micron filter.Splay was observed around the gate typical of a molding error.The discoloration is cosmetic and does not affect the functionality of the device.The reported complaint of condensation in the tubing could not be confirmed.The discoloration is typical of a molding error.The lot history of the provided labels were reviewed, no nonconformities were identified that may have contributed to the reported complaint.
|
| |
|