SYMMETRY SURGICAL INC SYMMETRY BOOKWALTER; HORIZONTAL FLEX BAR Back to Search Results
Catalog Number 50-4582
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the purchase date provided by the customer, the device has been in use for approximately 4 years.Additionally, the customer notified us that the device is being regularly lubricated.The instructions for use of this device has a warning stating that this device should not be lubricated as it could cause movement during use.Over many years of use and lubrications, the lubricant will begin to build up within the locking mechanism which could prevent it from locking properly.The customer was notified of the proper way to care for the device and it was recommended that the device be refurbished.(b)(4).Based on the above information, no further actions are required and this can be seen as the final report.If additional information is obtained that alleges any additional patient involvement or additional information pertinent to the investigation, a subsequent follow up report will be submitted.
 
Event or Problem Description
The customer alleged that the bookwalter horizontal flex bar became loose causing it to fall into the patient's abdomen.There was no harm to the patient.A delay in the procedure occurred to correct the placement of the devices.The length of the delay is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMMETRY BOOKWALTER
Common Device Name
HORIZONTAL FLEX BAR
Manufacturer (Section D)
SYMMETRY SURGICAL INC
3034 owen drive
antioch TN 37013
Manufacturer (Section G)
SYMMETRY SURGICAL INC
3034 owen drive
antioch TN 37013
Manufacturer Contact
brandi meath
3034 owen drive
antioch, TN 37013
6159645290
MDR Report Key17596456
Report Number3007208013-2023-00037
Device Sequence Number20466776
Product Code GAD
UDI-Device Identifier00887482018545
UDI-Public00887482018545
Combination Product (Y/N)N
Initial Reporter CountryMY
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number50-4582
Device Lot NumberR002S467/28
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2023
Initial Report FDA Received Date08/22/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-