| Model Number |
BML-V242QR-30 |
| Medical Device Problem Codes |
Break (1069); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The customer reported to olympus during a diagnostic endoscopic retrograde cholangiopancreatography (ercp), the single use mechanical lithotriptor had two failures and a modification to the device was required to complete the case.The flower basket handle was removed from the device in order to use the emergency lithocrush handle and the sheath had to be cut emergently.The procedure duration was around 4 hours which was very prolonged due to the product failure and anesthesia was also extended.The reported problem did not affect the outcome of the procedure.The procedure was completed with the emergency lithocrush.
|
| |
|
Event or Problem Description
|
|
This supplemental report is being created to include additional information received from the customer.When the original device was connected, no pressure was being applied to the stone and the handle retracted all the way out and then continued spinning.The handle was reattached with the same result leading the device to fail, therefore the rescue device was applied.Tension and pressure were applied in attempt to crush the stone, however the wire broke inside the rescue handle.The sheath surrounding the wire was then cut to modify the device and give more wire to wrap in the rescue handle.The emergency handle was able to start turning the wires and the stone was then crushed.
|
| |
|