| Catalog Number |
MM-410360 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported a right side rupture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
The device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data.Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture : observed, broken side assessed as unidentified (tear) and missing side assessed as inconclusive.(shell thickness was within specification).No additional observation.No further actions are required as no issues with the manufacturing process are observed.
|
| |
|