|
A territory manager reported an end user experienced an issue when using a picc from a bioflo picc (valved) 5fdl-55cm maximal barrier nursing kit.The 5f dual lumen picc was getting a tpn infusion and the hub cracked.The picc was removed and replaced.The patient did not experience any adverse effects or harm due to this incident.Tpn is ordered daily, administered nightly (either continuous or cyclic).Policy is to flush the line after use/completion of the bag, and there were no reports of tools used to tighten lines, all staff involved in discussion with review state they attached needleless connectors and tubing with finger tightening only.
|
|
Returned for evaluation was one (1) bioflo picc.As received, the female luer lock component of the white/purple valve was confirmed to be cracked.There were no other visual defects noted to the valve.Visual inspection of the type and location of the crack in the hub luer is indicative of the hub luer having an over-tightened connection, likely in the presence of tpn or other lipid based infusate.The customer's complaint description is confirmed for hub cracked.No component molding non-conformances or other damage was observed during visual inspection of the returned sample.The most likely root cause for the cracked female valve housing is due to an over tightened connection with a mating male luer (likely a needleless connector).A contributing factor to the over-tightened connection may be the mechanics/hand placement during infusion access.It is preferable to grasp the needleless connector (nc) when connecting a syringe/tubing to it rather than grasping the pasv valve luer hub while making a connection with the nc.Grasping the pasv valve hub while connecting a syringe/tubing set to the nc can transmit additional torque to the pasv female luer hub.Inadequate flushing of the device lumen may also be a contributing factor.The dhr for the packaging/assembly lots were reviewed for any deviations related to the reported hub crack failure mode.The review confirms that the lots met all material, assembly and performance specifications with no internal ncrs written.Labeling review: the dfu that is supplied in bioflo kits item number contains the following precaution: it is recommended that only luer lock accessories be used with the · bioflo¿ picc with endexo¿ and pasv¿ valve technology.Repeated over-tightening may reduce hub connector life.Do not use hemostats to secure or remove devices with luer lock hub connections.Flushing - recommended procedure: 1.Flush the catheter after every use, or at least every seven days when not in use, to maintain patency.Use a 10 ml syringe or larger.2.Flush the catheter with a minimum of 10 ml of sterile normal saline, using a "pulse" or "stop/start" technique.Warning: if using bacteriostatic saline, do not exceed 30 ml in a 24-hour period.3.Disconnect the syringe and attach a sterile end cap to each luer lock hub.Note: this is the recommended flush procedure for this catheter.If using a different procedure than listed above, the use of heparin may be necessary.Follow institutional protocol for catheter flushing.Precaution: incompatible drug delivery within the same lumen may cause precipitation.Ensure that the catheter lumen is flushed following each infusion.Precaution: if resistance is met when flushing, it is recommended that no further attempts be made.Further flushing may result in catheter rupture.Refer to institutional protocol for clearing occluded catheters.Precaution: place a cap on the hub after use to reduce the risk of contamination.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference(b)(4).
|