• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS BIOFLO PICC WITH ENDEXO AND PASV VALVE TECHNOLOGY; PERIPHERALLY INSERTED CENTRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS BIOFLO PICC WITH ENDEXO AND PASV VALVE TECHNOLOGY; PERIPHERALLY INSERTED CENTRAL CATHETER Back to Search Results
Catalog Number 75019
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of picc hub cracked cannot be confirmed, no sample was returned.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.The most likely root cause for the cracked female valve housing is due to an over tightened connection with a mating male luer (likely a needleless connector).A contributing factor to the over-tightened connection may be the mechanics/hand placement during infusion access.It is preferable to grasp the needleless connector (nc) when connecting a syringe/tubing to it rather than grasping the pasv valve luer hub while making a connection with the nc.Grasping the pasv valve hub while connecting a syringe/tubing set to the nc can transmit additional torque to the pasv female luer hub.Inadequate flushing of the device lumen may also be a contributing factor.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Labeling review: the dfu that is supplied in bioflo kits contains the following precaution: it is recommended that only luer lock accessories be used with the · bioflo¿ picc with endexo¿ and pasv¿ valve technology.Repeated over-tightening may reduce hub connector life.Do not use hemostats to secure or remove devices with luer lock hub connections.Flushing - recommended procedure: 1.Flush the catheter after every use, or at least every seven days when not in use, to maintain patency.Use a 10 ml syringe or larger.2.Flush the catheter with a minimum of 10 ml of sterile normal saline, using a "pulse" or "stop/start" technique.Warning: if using bacteriostatic saline, do not exceed 30 ml in a 24-hour period.3.Disconnect the syringe and attach a sterile end cap to each luer lock hub.Note: this is the recommended flush procedure for this catheter.If using a different procedure than listed above, the use of heparin may be necessary.Follow institutional protocol for catheter flushing.Precaution: incompatible drug delivery within the same lumen may cause precipitation.Ensure that the catheter lumen is flushed following each infusion.Precaution: if resistance is met when flushing, it is recommended that no further attempts be made.Further flushing may result in catheter rupture.Refer to institutional protocol for clearing occluded catheters.Precaution: place a cap on the hub after use to reduce the risk of contamination.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Event or Problem Description
Approximately 8 days post placement, the patient's 5f dl bioflo pasv picc was observed to have a leak coming from a hairline crack on the hub of the device during tpn infusion.As a result, the device was removed from the patient.The patient was receiving daily dose of tpn.The device was replaced with a new of the same.Of note, staff remembers he asked to be disconnected and reconnected more often than most patients, suspect this highlights susceptibility to cracking with repetitive placement/removal of connectors.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO PICC WITH ENDEXO AND PASV VALVE TECHNOLOGY
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key17597841
Report Number1317056-2023-00115
Device Sequence Number20725453
Product Code LJS
UDI-Device IdentifierH965750191
UDI-PublicH965750191
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K121089
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Catalogue Number75019
Device Lot Number5776478
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2023
Initial Report FDA Received Date08/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
Patient Weight58 KG
-
-