Model Number 1010
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hypersensitivity/Allergic reaction (1907)
Date of Event 03/14/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) was part of a system revision due to suspected nickel allergy.There were no additional adverse patient effects reported.This s-icd was explanted.
 
Event or Problem Description
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) was part of a system revision due to suspected nickel allergy.There were no additional adverse patient effects reported.This s-icd was explanted.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, additional information from product investigation indicates that the allegation against the device was not confirmed.This device was analyzed, passed all the baseline tests and exhibited normal device functions.This supplemental is being filed to capture the investigation results.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SQ-RX PULSE GENERATOR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17599387
Report Number2124215-2023-45282
Device Sequence Number15346006
Product Code LWS
UDI-Device Identifier00802526562808
UDI-Public00802526562808
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P110042/S000
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date11/17/2015
Device Model Number1010
Device Catalogue Number1010
Device Lot NumberA015991
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/08/2023
Supplement Date Received by Manufacturer08/22/2023
Initial Report FDA Received Date08/22/2023
Supplement Report FDA Received Date08/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age59 YR
Patient SexFemale
-
-