• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. 145-DEG PE 38MM CONST HUM LINER +2.5; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. 145-DEG PE 38MM CONST HUM LINER +2.5; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-38-13
Medical Device Problem Codes Detachment of Device or Device Component (2907); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 08/10/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via clinical study, that approximately 6 months postop the initial implant, this 60 yo male patient was revised due to disassociation of polyethylene.The patient reached out to the side while exercising, felt a "clunk".The patient¿s outcome was last known as resolved.Devices will not be returned.
 
Additional Manufacturer Narrative
D10.Concomitants: a239027 300-01-13 - equinoxe, humeral stem primary, press fit 13mm, a185543 320-08-38 - glenosphere exp 38mm +4mm offset, 5735293 320-15-06 - rs glenoid plate ext cag +10mm cage peg, a300200 320-10-00 - equinoxe reverse tray adapter plate tray +0.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: d1, h6.Mdr section codes updated/corrected: a, b, c, d, e, g.The revision reported was likely the result of disassembly between the humeral liner and humeral tray.However, this cannot be confirmed and potential contributions from user conditions, an unreported traumatic event, incomplete seating of the humeral liner during implantation, forceful contact between the liner and glenoid bone (scapular notching), or a combination of the above to the reported disassembly of the humeral liner cannot be determined as the devices were not returned for evaluation and images/radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
145-DEG PE 38MM CONST HUM LINER +2.5
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct.
gainesville FL 32563
Manufacturer Contact
kate jacobson
MDR Report Key17599600
Report Number1038671-2023-02018
Device Sequence Number20443084
Product Code KWT
UDI-Device Identifier10885862086686
UDI-Public10885862086686
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number320-38-13
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2023
Supplement Date Received by Manufacturer09/16/2025
Initial Report FDA Received Date08/22/2023
Supplement Report FDA Received Date09/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age60 YR
Patient SexMale
Patient Weight83 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-