EXACTECH, INC. 145-DEG PE 38MM CONST HUM LINER +2.5; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Catalog Number |
320-38-13 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
08/10/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported via clinical study, that approximately 6 months postop the initial implant, this 60 yo male patient was revised due to disassociation of polyethylene.The patient reached out to the side while exercising, felt a "clunk".The patient¿s outcome was last known as resolved.Devices will not be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
D10.Concomitants: a239027 300-01-13 - equinoxe, humeral stem primary, press fit 13mm, a185543 320-08-38 - glenosphere exp 38mm +4mm offset, 5735293 320-15-06 - rs glenoid plate ext cag +10mm cage peg, a300200 320-10-00 - equinoxe reverse tray adapter plate tray +0.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: d1, h6.Mdr section codes updated/corrected: a, b, c, d, e, g.The revision reported was likely the result of disassembly between the humeral liner and humeral tray.However, this cannot be confirmed and potential contributions from user conditions, an unreported traumatic event, incomplete seating of the humeral liner during implantation, forceful contact between the liner and glenoid bone (scapular notching), or a combination of the above to the reported disassembly of the humeral liner cannot be determined as the devices were not returned for evaluation and images/radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|