| Catalog Number |
68LP-250 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/30/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported, right side deflation.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, corrected and/or changed data: b.5, d.6b, h.6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: observed 1 opening assessed as unidentified (tear) opening (shell thickness was within specification).No additional observations are performed.No further actions are required as no manufacturing issues are observed.
|
| |
|
Search Alerts/Recalls
|