| Model Number |
VTICM5_13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2; -9.5/+3.0/122 (sphere/cylinder/axis) implantable collamer lens into the patient's left eye (os) on (b)(6) 2023.The patient experienced low vaulting with no lens rotation.Lens remains implanted.The cause of the event was reported as unknown.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - the reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2 implantable collamer lens of diopter -9.5/3.0/122 (sphere/cylinder/axis) into the patient's left eye (os) on (b)(6) 2023.The lens was exchanged with a longer length lens on (b)(6) 2023 due to low vault without rotation.This resolved the problem.The cause of the event is reported as unknown.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - health effect impact code 4629: lens was exchanged should be added to supplemental mdr #1.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: the lens was returned dry in a microcentrifuge vial.Visual inspection found haptic torn and foreign material on the lens surface, specifically, residue on the lens.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
G3 - 20-dec-2023 corrected to 28-nov-2023.Claim # (b)(4).
|
| |
|