• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W&H DENTALWERK BUERMOOS GMBH MAXIMA ELITE; HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W&H DENTALWERK BUERMOOS GMBH MAXIMA ELITE; HANDPIECE Back to Search Results
Model Number MAXIMA ELITE 2 1:5
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 06/26/2023
Type of Reportable Event Serious Injury
Event or Problem Description
After several uses, the product heated up and burned three (3) patients.
 
Additional Manufacturer Narrative
Related report number: (b)(4).The complained contra-angle handpiece was examined by w&h and no technical fault on the part of w&h could be detected.Thus, the initial assessment (lack of maintenance) of the distributor as the cause of the strong heating can be confirmed as plausible.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXIMA ELITE
Common Device Name
HANDPIECE
Manufacturer (Section D)
W&H DENTALWERK BUERMOOS GMBH
ignaz-glaser-strasse 53
buermoos, 5111
AU  5111
Manufacturer (Section G)
W&H DENTALWERK BUERMOOS GMBH
ignaz-glaser-strasse 53
buermoos, 5111
AU   5111
Manufacturer Contact
gerhard weidler
ignaz glaser strasse 53
buermoos, 5111
AU   5111
MDR Report Key17600814
Report Number3011300255-2023-00005
Device Sequence Number17719053
Product Code EFB
UDI-Device Identifier00304040153857
UDI-Public0100304040153857
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K162926
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMAXIMA ELITE 2 1:5
Device Catalogue Number30431000
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2023
Initial Report FDA Received Date08/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-