The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd module with drive pcb.In addition, our technician confirmed that the angle wire fluid damage, the insertion flexible tube fluid damage, the segment fluid damage, the control body fluid damage, the remote control wire fluid damage, the lcb (light carrying bundle) fluid damage, the light guide cable fluid damage, the lg connector fluid damage, the rubber trim collar broken, the lcb (light carrying bundle) broken, and the root brace rubber (insertion flexible tube) cut; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|