STRYKER ENDOSCOPY-SAN JOSE AIM 4K COUPLER, 20MM, C-MOUNT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 1688020122
Medical Device Problem Codes Fogging (1253); Moisture or Humidity Problem (2986)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was foggy image.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.Alleged failure: foggy image.Probable root cause:.¿ cables.¿ hdmi cables.¿ connectors.¿ digital board.¿ power supply.¿ filter / fuse.¿ ac inlet board.¿ main board.¿ transition board.¿ intermittence between transition board and ch connector.¿ software.¿ camera head (sensor, sensor cable).¿ coupler (dust on optics, scratched optics, broken / shifted lenses, focusing mechanism, endoclamp, zoom ring).¿ problem toggling light source.¿ connected monitors.¿ leaking.¿ poor grounding (e.G camera head connection from cable to transition board.).¿ no/incorrect communication with light source.¿ soaking cap disconnection during sterilization.¿ electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge.¿ cybersecurity attack.¿ pendulum ch inertial measurement unit (imu).¿ incident light from surgical scene is reflected by coupler/ch window.¿ use error.The reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that there was foggy image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIM 4K COUPLER, 20MM, C-MOUNT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key17602718
Report Number0002936485-2023-00851
Device Sequence Number6206335
Product Code GCJ
UDI-Device Identifier07613327382389
UDI-Public07613327382389
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K182160
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 07/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1688020122
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/02/2023
Supplement Date Received by Manufacturer08/02/2023
Initial Report FDA Received Date08/23/2023
Supplement Report FDA Received Date07/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-