| Model Number |
SM300 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Through follow-up with the user facility, it was disclosed that chloraprep solution was used on the patient's skin as a prep solution.The chloraprep solution contains a high percentage of isopropyl alcohol which can be used to remove the skin marker as stated in the instructions for use.The surgical skin marker instructions for use include the following language, "skin surface must be clean and dried before marking skin.Mark the desired areas on intact skin with the skin marker according to facility protocol.Allow the markings to dry thoroughly before prepping.Prepare and apply prep solution as instructed by prep solution manufacturer.Markings may fade slightly but remain legible.Remove markings with isopropyl alcohol if desired.Markings will disappear naturally with bathing in a few days." it is likely that the use of the chloraprep solution allowed the skin marker to rub off of the patient's skin.The user facility was provided with a copy of the instructions for use.The device history record was reviewed, and no abnormalities were noted.There have been no other complaints associated with this lot.A 3-year complaint review indicates this to be an isolated event.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The user facility reported that at the beginning of a patient procedure, it was noted that the ink from their skin marker that was used to draw on the patient prior to the procedure had rubbed off of the patient's skin.The markings had to be redrawn resulting in a procedure delay.The procedure was completed successfully.No report of injury.
|
| |
|
Search Alerts/Recalls
|