Model Number 866148
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Phone number provided: (b)(6).Philips is in process of obtaining additional information.A final report will be submitted upon completion of the investigation.
 
Event or Problem Description
The customer reported that icca has an intake miscalculation issue.The device was in use on a patient.There was no report of patient or user harm.
 
Additional Manufacturer Narrative
A functional analysis was performed and identified icca had an intake miscalculation issue.The remote service engineer (rse) stated was miscalculation and was known issue in past year and r&d had released fco in enable totals watchdog service, in that site watchdog service was enabled and values updating by db services was expected one.Rse observed drain out values were updated backwards from 120ml to 40ml and 80ml to 30ml , when that had happened next time, kindly did not modified the value and reported to mrc.Rse didn¿t have clear answer at that moment on root cause for that issue, could be glitch during the job run, no evidences recorded in event logs.Watchdog services runs every 480 minutes, the issue was fixed by next job run at 00:10 of watchdog job in database.A review of the risk management file indicates the problem reported by the customer is a low potential severity which would not reasonably cause or contribute to death or serious injury if the problem were to reoccur.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLISPACE CC & ANESTHESIA
Common Device Name
INTELLISPACE CC & ANESTHESIA
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key17603210
Report Number1218950-2023-00596
Device Sequence Number6308182
Product Code DXJ
UDI-Device Identifier00884838085565
UDI-Public00884838085565
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K100272
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number866148
Device Catalogue Number866148
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2023
Supplement Date Received by Manufacturer06/19/2023
Initial Report FDA Received Date08/23/2023
Supplement Report FDA Received Date09/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-