| Catalog Number |
03P85-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Adult Respiratory Distress Syndrome (1696)
|
| Date of Event |
08/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2023, abbott point of care was contacted by a customer regarding i-stat cg4+ cartridge that yielded a suspected discrepant ph results on a 63 year old female patient with respiratory distress.The patient was admitted and still hospitalized.There was no additional patient information available at the time of this report.Return product is not available for investigation.Method: i-stat.Date: (b)(6) 2023, collected: 15:23, 15:23.Tested: 7.233.Result: arterial.Sample: (sample a).Method: i-stat.Date: (b)(6) 2023, collected: 15:30, 15:30.Tested: 7.142.Result: arterial.Sample: (sample b).Method: i-stat.Date: (b)(6) 2023, collected: 15:38, 15:38.Tested: 7.410.Result: arterial.Sample: (sample c).There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 01-nov-2023.A review of the device history record (dhr) confirmed the cartridge lot met finished goods release criteria.Retained cartridge testing met the acceptance criteria found in q04.01.003 rev.Al, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been determined for cg4+ cartridge lot m23095.
|
| |
|
Search Alerts/Recalls
|