| Catalog Number |
S-55-150-120-P6 |
| Medical Device Problem Codes |
Material Separation (1562); Device Dislodged or Dislocated (2923); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a moderately calcified right superior femoral artery (sfa).Atherectomy was performed to prep the lesion.The 5.5x150mm supera was advanced without issue and thought to be fully deployed at the intended site.However, upon removal right when the delivery system was out it was noted that the stent came along with it and the stent came off landed on the patient along with the separated nosecone.Nothing was left in the patient as this occurred outside the anatomy.The device was removed under fluoroscopy.A new supera was used to successfully complete the procedure.There was no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported material separation was able to be confirmed.The reported activation failure and the reported device dislodged or dislocated - stent were unable to be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that after stent deployment during device removal the tip inadvertently got caught with the deployed stent resulting in the reported activation/deployment failure.Manipulation of the compromised device during removal resulted in the reported stent dislodgement, the noted flattened/twisted sheath and ultimately resulted in the reported tip separation/noted inner member and tip jacket separations.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Search Alerts/Recalls
|