| Model Number |
50-347-07-61 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
| Date of Event |
07/26/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that screws securing a distractor became loose from bone.The physician removed the screws and replaced them with the same size and model screws while leaving the distractor in place.The screws were noted to be intact and undamaged upon explant.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at kls se.During the investigation the product lot number identified was reviewed in the device history records.The dhr review showed no discrepancies or anomalies.The failure root cause cannot be determined due to no device being returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.Corrections: d3 establishment name - kls martin se & co.Kg.G1 contact office establishment name - kls-martin l.P.G1 contact office manufacturing site name and address - kls martin se & co.Kg.Kolbinger strasse 10.Muehlheim/donau, deu 78570, deu.
|
| |
|