| Model Number |
15 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and was unable to duplicate the reported event.After replacing the system pcb assembly, proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
A customer contacted stryker to report that their device would not complete its boot up cycle when powering on.As a result, defibrillation therapy would not be available, if needed.There was no patient use associated with the reported event.
|
| |
|