Visual inspection: during the investigation, no significant deformations or damage of the valves was determined.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the valves operate in their respective relevant positions, within the permissible tolerance.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection: after dismantling of the valves, no deposits were found inside.To visualize possible invisible proteins/deposits in the shunt system, the valves were stained with a staining solution.Results: based on our investigation results, we cannot determine any functional deviation in the product.At the time of the examination, it is not clear to us how the aforementioned functional impairment occurred.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|
|
It was reported that a progav 2.0 shuntsystem (part # fx596t) was implanted during a procedure performed on (b)(6) 2023.According to the complainant, the shunt showed an under-drainage.The patient underwent a revision procedure performed on (b)(6) 2023.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 4 months, weight: 4.9 kilograms (kg), height: 57 centimeters (cm), gender: female.
|