ETHICON INC. VCL+ UD 18IN 5-0 S/A P-1 PRM; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCP490G
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2023 and suture was used.During the procedure, when using the suture to close an incision, the needle became detached from the suture while pulling it.According to nurse in charge, there was no harm to the patient and the surgery ended successfully.There were no adverse patient consequences.Additional information was requested.
 
Additional Manufacturer Narrative
(b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.¿ trade name - irgacare®.¿ active ingredient(s) ¿ triclosan.¿ dosage form ¿ suture/solid/parenteral.¿ strength ¿ = 472 g/m.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H6 component code: g07002 device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: was there any adverse consequence associated with the patient? ¿ according to the nurse in charge there were no adverse consequence associated with the patient.Note: related events reported via 2210968-2023-06139.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 9/12/2023.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 pending evaluation of returned device a device has been received, however clarification is requested whether the product belongs to this complaint.The following additional information was requested: 1.Could you please clarify if the additional device belong to this complaint? 1a.If yes, did a device malfunction occur during the same procedure? 2.If not, could you please clarify if the additional received product belongs to a different complaint? if so, can you please provide the complaint number.3.If the device was not reported before, please provide more details of device malfunction.4.If the product doesn¿t belong to any complaint, please let us know so we can discard it or advise if the product is required to be returned.5.If you have the correct complaint product in your possession, please forward the product to ethicon as soon as possible.If you have already sent the product to ethicon, please reply to ethwcqcom@its.Jnj.Com with the ups, dhl or fedex tracker number.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information was requested and the following was obtained: the additional products received are not related to this complaint or any complaint we know of.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Investigation summary :the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.The returned sample determined that it was received, two unopened samples that pertain to the product code vcp490g.In order to evaluate the condition of the returned samples, the packets were opened.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 18IN 5-0 S/A P-1 PRM
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
col salvacar
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key17611491
Report Number2210968-2023-06140
Device Sequence Number15131658
Product Code GAM
UDI-Device Identifier10705031053632
UDI-Public10705031053632
Combination Product (Y/N)Y
Initial Reporter CountryIS
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCP490G
Device Lot NumberSJMTCD
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2023
Supplement Date Received by Manufacturer08/24/2023
09/20/2023
10/23/2023
Initial Report FDA Received Date08/24/2023
Supplement Report FDA Received Date09/12/2023
10/18/2023
11/07/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-