| Model Number |
AR-1926BC |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 7/27/2023, it was reported by an arthrex employee via email that (2) ar-1922bc biocomposite pushlock anchor pulled out.This was discovered when the surgeon tried to fix the lateral row with the pushlock anchor during an rcr procedure on (b)(6) 2023.Additional information requested.
|
| |
|
Event or Problem Description
|
|
On 7/27/2023, it was reported by an arthrex employee via email that (2) ar-1922bc biocomposite pushlock anchor pulled out.This was discovered when the surgeon tried to fix the lateral row with the pushlock anchor during an rcr procedure on (b)(6) 2023.Additional information requested.Part number was updated 08/24/2023 to (qty.2) ar-1926bc biocomposite pushlock, as this was the complaint device and not (2) ar-1922bc.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint is not confirmed as neither the anchor nor the eyelet was returned for investigation.One unpacked ar-1926bc was received for investigation without the anchor and eyelet.No problem found for the reported failure as the anchor and eyelet were not returned and no photos were provided.Visual evaluation noted that the inner shaft was bent at the distal end at the tip.No other issues were found.No functional testing performed due to the damage of the device.The method of bone preparation employed during the procedure, as well as the bone quality encountered, was not provided.The most likely cause(s) for this failure can be attributed to user error due to improper bone preparation, misaligned insertion, and/or prying/levering the device during insertion.According to dfu-0087 section e.Warnings.6.Bioabsorbable only: attempting implantation into hard, dense bone and/or drilling/punching smaller diameter holes than recommended may cause failure (breakage) of the implant during insertion.Section g.Precautions.3.Make sure to use the recommended drill bit or punch to create the bone socket.4.Pushlock and swivelock suture anchors only: during anchor insertion, ensure that the angle of anchor insertion is coaxial to that of the previously prepared bone socket.5.Pushlock and swivelock suture anchors only: insert the driver into the bone socket until the anchor body makes contact with the bone.Preview and adjust suture tension, if necessary.Tension will not increase during final advancement of the anchor body.6.Pushlock and swivelock suture anchors only: ensure that the anchor body is in full contact with the bone before advancing the anchor body into the prepared bone socket.Refer to the investigation pictures.
|
| |
|