C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA 4F DL MBPLUS (3CG, GUARDIVA, PROBE COVER); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and is pending evaluation.Results are expected soon.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer "nurse was placing a picc yesterday and then when she was flushing the ports and got to the purple one, found it was leaking at the hub.She did an over the wire exchange." response received on 11-aug-2023."event did not involve an urgent/life threatening medical situation or cause clinically significant delay.Since my nurse was able to do an over-the-wire exchange, there were no complications to the patient or significant delay of care.The catheter had not yet been secured at the time the defect was noticed.".
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leaking catheter was confirmed.The product returned for evaluation was one 4 fr d/l powerpicc provena.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue occurs due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated and a split was observed just distal of the molded joint.The catheter split contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: ¿ damage which was circumferentially aligned.¿ fracture edges which were rounded and polished due to repeated material wear.¿ overall elliptical shape to the fracture cross-section (a result of repeated kinking of the tubing).An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.The damage location suggested that catheter securement, access and maintenance techniques may have contributed.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer "nurse was placing a picc yesterday and then when she was flushing the ports and got to the purple one, found it was leaking at the hub.She did an over the wire exchange." response received on 11-aug-2023."event did not involve an urgent/life threatening medical situation or cause clinically significant delay.Since my nurse was able to do an over-the-wire exchange, there were no complications to the patient or significant delay of care.The catheter had not yet been secured at the time the defect was noticed.".
|
| |
|