| Medical Device Problem Codes |
Material Fragmentation (1261); Material Separation (1562)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/11/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Other, other text: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient was scheduled for port a cath removal "dt" non-functioning port.An x-ray was taken prior to incision and a severed port was noted.Partial port removed.There was clinician injury and medical intervention.Patient required additional care.Adverse patient effects are unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: d4: udi updated - (b)(4).H6: evaluation codes updated device evaluation: one sample was returned.During the visual inspection it was observed that the catheter connected to the port was broken.No damage, cuts or other defects were detected.Complaint was not able to be confirmed.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.
|
| |
|