ST PAUL PORT-A-CATH IMPLANTABLE ACCESS SYSTEM; PORT AND CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAV Back to Search Results
Catalog Number 21-4455-24
Medical Device Problem Codes Break (1069); Material Separation (1562)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/13/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient was sent to have the device removed due to it breaking.The device was implanted (b)(6) 2023 and recently developed an "issue".Adverse patient effects are unknown.
 
Additional Manufacturer Narrative
Other text: d4: lot number, expiration date and h4: device manufacture date is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
Other, other text: h6: evaluation codes updated.Device evaluation: two samples were received for investigation.After examining under magnification, it appears that the catheters failed in tension on one side of the catheter at the end of the outlet due to bending.On one of the two, the tear propagated across the catheter causing complete separation.A dhr review was unable to be completed as the lot number was not provided.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORT-A-CATH IMPLANTABLE ACCESS SYSTEM
Common Device Name
PORT AND CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAV
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17616190
Report Number3012307300-2023-08390
Device Sequence Number2608822
Product Code LJT
UDI-Device Identifier10610586032707
UDI-Public10610586032707
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K072657
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number21-4455-24
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2023
Supplement Date Received by Manufacturer04/01/2024
Initial Report FDA Received Date08/24/2023
Supplement Report FDA Received Date06/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-