| Catalog Number |
21-4455-24 |
| Medical Device Problem Codes |
Break (1069); Material Separation (1562)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient was sent to have the device removed due to it breaking.The device was implanted (b)(6) 2023 and recently developed an "issue".Adverse patient effects are unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: d4: lot number, expiration date and h4: device manufacture date is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: h6: evaluation codes updated.Device evaluation: two samples were received for investigation.After examining under magnification, it appears that the catheters failed in tension on one side of the catheter at the end of the outlet due to bending.On one of the two, the tear propagated across the catheter causing complete separation.A dhr review was unable to be completed as the lot number was not provided.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.
|
| |
|