GYRUS ACMI, INC. EZDILATE BALLOON DILATOR (FW) 18-19-20; OESOPHAGEAL BALLOON CATHETER Back to Search Results
Model Number BD-400P-2080
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H4 device manufacturer date: date of manufacture cannot be determined since the lot number was not provided.The subject device has not been returned to olympus for further evaluation and testing.The investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
 
Event or Problem Description
The customer reported that the balloon popped during a therapeutic esophageal dilation procedure.The intended procedure was completed using another device.There was no report of patient or user injury due to the event.
 
Additional Manufacturer Narrative
Correction: the lot number was inadvertently omitted from the initial report.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The investigation concluded that this nonconformance observed in this complaint was categorized as a process defect.An exhaustive revision of the process was completed, founding that occlusion partial or total was generated in the proximal bonding station due to misalignment of the d-mandrel or mandrel during the preparation for bonding process.The operator has not enough visibility to ensure the correct position of the d-mandrel or mandrel, causing melting material smearing occluding partial or total the diameter of the extrude.Causing two events, the first is the diameter is completely occluded which would prevent the balloon from inflating.And the second event would be that the diameter is partially occluded which would cause it to take longer to inflate or deflate.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZDILATE BALLOON DILATOR (FW) 18-19-20
Common Device Name
OESOPHAGEAL BALLOON CATHETER
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17619815
Report Number3011050570-2023-00126
Device Sequence Number20679073
Product Code KNQ
UDI-Device Identifier00821925033214
UDI-Public00821925033214
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT-KNQ
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBD-400P-2080
Device Lot Number383361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2023
Supplement Date Received by Manufacturer11/13/2023
Initial Report FDA Received Date08/24/2023
Supplement Report FDA Received Date11/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/14/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MAJ-1740 EZDILATE¿ INFLATION DEVICE.
Patient SexUnknown
-
-