• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number ATG120144
Medical Device Problem Codes Deflation Problem (1149); Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/29/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 11/2025).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during angioplasty procedure in the central vein, the pta balloon allegedly failed to deflate.It was further reported that after the dilation, the balloon allegedly detached from the catheter.It was also reported that the pta balloon catheter was removed using snare and withdrawn the punctured balloon into the sheath and completed the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.Two photos were reviewed.The first photo shows a completely detached and inverted balloon.The balloon was bloody.The second photo shows the detached balloon and catheter.No other anomalies are noted in the submitted photos.No objective evidence of a puncture site nor tear on the balloon was noted as reported in the user's condition and due to the device's condition.Based on the submitted photos, the balloon shows the evidence of complete detachment from the catheter and the balloon was also noted to be fully inverted within itself.However, the photos show no objective signs of puncture on the balloon material as reported by the user.Hence, the investigation into the reported balloon detachment and difficulty into removing the balloon is confirmed and the investigation remains inconclusive for the reported deflation issue.A definitive root cause for the reported balloon detachment, deflation issue and difficulty into removing the balloon could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 11/2025), g3, h6 (method).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during angioplasty procedure in the central vein, the pta balloon allegedly failed to deflate.It was further reported that after the dilation, the balloon allegedly detached from the catheter.It was also reported that the pta balloon catheter was removed using snare and withdrawn the punctured balloon into the sheath and completed the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17620612
Report Number2020394-2023-00620
Device Sequence Number20662581
Product Code DQY
UDI-Device Identifier00801741060762
UDI-Public(01)00801741060762
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K181323
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberATG120144
Device Lot Number93YG0162
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/31/2023
Supplement Date Received by Manufacturer11/15/2023
Initial Report FDA Received Date08/24/2023
Supplement Report FDA Received Date11/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-