| Model Number |
PV-ACREC-04M |
| Medical Device Problem Codes |
Material Discolored (1170); Insufficient Device Problem Information (3190); Unexpected Color (4055)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is in process.The device will not be returned for evaluation.When the investigation is complete, a supplemental mdr will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that a pelvic implant was observed to be discolored post-sterilization at the hospital for a surgery on (b)(6) 2023.It was reported that the metal implant was sterilized with the plastic components.The device was implanted in the patient during this procedure.No further information has been made available.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the pelvic implant and anatomic models for a my3d pelvic reconstruction case on (b)(6) 2023 were sterilized together.Discoloration was reported to be seen on the implant post sterilization.The procedure was able to be completed successfully.No surgical delay was caused by this issue.Visual, dimensional, functional and material analysis could not be conducted as the part was implanted.Production records were reviewed, all manufacturing and associated inspections were completed successfully.Sterilization batch record review was not performed as these implants and accessories are provided nonsterile.These parts undergo sterilization at the hospital prior to surgery.The my3d personalized pelvic reconstruction ifu specifies that the metal implant and plastic accessories should not be sterilized together, which likely contributed to this issue.
|
| |
|
Search Alerts/Recalls
|