CAREFUSION, INC VITAL ADVANCED WIRE CUTTER 7IN RIGHT; CUTTER, WIRE Back to Search Results
Catalog Number OS3032
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/14/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pr # (b)(4).18aug2023 writer sent the customer an email acknowledging receipt of the complaint and requested follow up information including as to if there was any patient impact related to this event.Writer provided contact information.Device not yet evaluated, if the device is evaluated a follow up will be sent.
 
Event or Problem Description
"black insert piece cracked in half during procedure".No patient impact.22aug2023 additional information: what is the lot number for the defective cutter? k22xbf.How many occurrences of the defective cutter(s)? one.What was the procedure that was being performed? cabg.Did any part the instrument fall into the patient¿s body, and if so how was it retrieved? yes, one broken piece retrieved by surgical team.Was there a medical procedure performed to verify if the instrument was in the patient¿s body, such as an x-ray? no.What was the patient¿s outcome? no harm.Was the procedure completed as planned? yes.No further information is not available.
 
Event or Problem Description
"black insert piece cracked in half during procedure".No patient impact.22aug2023 additional information: what is the lot number for the defective cutter? k22xbf.How many occurrences of the defective cutter(s)? one.What was the procedure that was being performed? cabg.Did any part the instrument fall into the patient¿s body, and if so how was it retrieved? yes, one broken piece retrieved by surgical team.Was there a medical procedure performed to verify if the instrument was in the patient¿s body, such as an x-ray? no.What was the patient¿s outcome? no harm.Was the procedure completed as planned? yes.No further information is not available.
 
Additional Manufacturer Narrative
Pr # (b)(4) follow up: the sample was provided, and an evaluation was performed.The root cause is misuse of the instrument.The reason for the crack in the rubber guard of the wire cutter is that a kirschner wire was cut twice in succession with minimal clearance.During the second cut, the sharp end of the wire was pressed into the splash guard, which then naturally snapped.The function of the rubber guard is no longer guaranteed by the crack.Use of a device for a task other than what it is intended for will usually result in a damaged or broken device.Prior to use, inspect devices to ensure proper function and condition.Do not use devices if they do not satisfactorily perform their intended function or if they have physical damage.There have been no issues identified with the material or manufacturing process.The product would have been manufactured and tested according to the specifications.A review of the device history records (dhr) did not identify and issues that would have contributed to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITAL ADVANCED WIRE CUTTER 7IN RIGHT
Common Device Name
CUTTER, WIRE
Manufacturer (Section D)
CAREFUSION, INC
75 north fairway drive
vernon hills IL 60061
Manufacturer (Section G)
CAREFUSION, INC
75 north fairway drive
vernon hills IL 60061
Manufacturer Contact
anna wehrheim
75 north fairway drive
vernon hills, IL 60061
8015652341
MDR Report Key17625146
Report Number1423507-2023-00098
Device Sequence Number15274627
Product Code HXZ
UDI-Device Identifier10885403102769
UDI-Public(01)10885403102769
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/25/2023
Report Date (Section F) 08/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberOS3032
Device Lot NumberK22XBF
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Device Age9 MO
Event Location Hospital
Date Report to Manufacturer08/25/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/14/2023
Supplement Date Received by Manufacturer08/14/2023
Initial Report FDA Received Date08/25/2023
Supplement Report FDA Received Date10/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-